Skip to main content

PAF 113 Test System

PAF 113 Test System for both particle separation and gas absorption/desorption of of air filters for passenger compartments (DIN 71460-1/2, ISO 11155-1/2)

The PAF 113 test system combined the advantages of PAF 111 particle test system and PAF 112 gas testing system. It povides a maximum of paramater testing for cabin air filters.
Logo
Logo

Standards

DIN 71460-1
DIN 71460-2
ISO 11155-1
ISO 11155-2
ISO 10121-1
ISO 10121-2
Interested?We will gladly advise you

+49 (351) 2166 430

contact form

Benefits

  • standard compliant testing of filter media and filter elements
  • combined testing with particles and with gases
  • very flexible system configuration possible
Parameter titleUnitValue
test flow rate m³/h 70 ... 700
test aerosol - DEHS, DOP, PAO, KCl, NaCl, ...
test gases - n-Butane 80 ppm, toluene 80 ppm, sulfur dioxide 30 ppm, hydrogen sulfide 10 ppm
measuring range, differential pressure Pa 0 ... 1000 (± 2 %)
dimension of test specimen mm max. 600 × 300
operating medium, gas/air - room air at standard test conditions (23°C ± 2 K; 50 % ± 2 % RH)
power supply - 400 VAC / 3P / 75 A / 50 ... 60 Hz and 3 × 230 VAC / 2 P / 16 A / 60 Hz
aerosol generation technology - dust disperser SAG 410/L, Atomizer ATM 220, ...
aerosol conditioning technology - Electrostatic Aerosol Neutraliser EAN 581
aerosol measurement instruments - aerosol spectrometer LAP 322
gas analysers - FID, NDIR
sensors, others - temperature, relative humidity, absolut pressure
dimensions (w × h × d) mm 3700 × 3000 × 3700

Silimar products

PAF 111

PAF 111 test system for testing particle separation of air filters for passenger compartments (DIN 71460-1, ISO/TS 11155-1)

Cabin Air Filter Test System PAF 111, front view
detail view

PAF 112

The PAF 112 Filter Test System for gas adsorption/desorption of air filters for passenger compartments (DIN 71460-2, ISO 11155-2)

Cabin Air Filter Test System PAF 112, frontal view